Tribological evaluation.pdf

(1401 KB) Pobierz
PII: S0142-9612(96)00152-4
ELSEVIER
PI1 SO142-9612
(96)
00152-4
Eiomoteriols 18 (1997) 441-447
0 1997 Elsevier Science Limited
Printed in Great Britain. All rights reserved
014%9612/97/$17.00
Tribological evaluation of oxidized
zirconium using an articular cartilage
counterface: a novel material for
potential use in hemiarthroplasty
A.M. Pate1 and M. Spector
Department of Orthopedic Surgery, Brigham and Women’s Hospital, Harvard Medical School, Boston, MA 02115,
USA; and Rehabilitation Engineering R&D Laboratory, BrocktonlWest Roxbury VA Medical Center, West Roxbury,
MA. USA
Problems with total joint replacement that have surfaced in recent years have made reconsideration of
more conservative hip reconstructive procedures compelling. Moreover, procedures such as hemiar-
throplasty might benefit from newly developed materials that could provide more favourable tribolo-
gical performance when employed as a counterface for articulation with articular cartilage. The
objective of this study was to evaluate the tribology of a new biomaterial, oxidized zirconium, with
articular cartilage in a laboratory test apparatus. Oxidized zirconium components are produced by
oxidizing the zirconium alloy to form a relatively thick (7,um), adherent, abrasion-resistant ceramic
surface. We found that the coefficient of friction of bovine articular cartilage rubbed against oxidized
zirconium was lower than with cobalt-chromium alloy control surfaces, and that there was a trend
toward less wear with oxidized zirconium. A defined layer of degraded tissue was found on the
surface of the articular cartilage specimens, providing some indication of the mechanism of wear.
Results of this study warrant further investigation
of oxidized zirconium as the bearing surface for
hemiarthroplasty.
0 1997 Elsevier Science Limited. All rights reserved
Keywords: Hemiarthroplasty,
wear, articular cartilage, oxidized zirconium
Received 19 June 1996; accepted 3 September 1996
Concern about the longevity of total joint replacement
has renewed interest in hemiarthroplasty as a treatment
for certain hip disorders. Hemiarthroplasty has been
well established as a practical and beneficial treatment
for elderly patients with femoral neck fractures
because of low morbidity and expedient return to
independence14. Young patients with avascular
necrosis of the femoral head have also been considered
to be good candidates for conservative reconstructive
surgical procedures involving hemiarthroplasty.
However, reports of relatively early failure19234* have
made this indication for the prosthesis questionable.
Causes of failure of hemiarthroplasty prostheses
include pain and the erosion (i.e. wear) of acetabular
articular cartilage and bone. These processes are
probably related in that wear debris can elicit
inflammatory processes that provoke the release of
neurotransmitters associated with pain. Moreover,
innervation of acetabular tissues with nociceptive
nerve fibres7 can be accompanied by an increase in the
amount of released neurotransmitters (e.g., substance
P) that are known to promote inflammations-10
and
degradative processes that might facilitate erosion of
tissue by the prosthesis. This might be reflected, in
part, in the high incidence of protrusion of the hip in
rheumatoid arthritis. Questions remain regarding the
relationship between design features and the erosive
processes underlying failure of hemiarthroplasty.
Design specifications for hemiarthroplasty prostheses
include head size, stem design, method of fixation (i.e.
cemented or non-cemented), unipolar versus bipolar
articulation and material from which the head is made.
Initially, prostheses were available in few head sizes,
leading to incongruence that resulted in higher contact
stresses. In addition, the non-cemented stem designs
were not as canal filling as was later found to be
necessary for mechanical stability. Bipolar devices
were developed to reduce the sliding contact with
tissue. However, despite the fact that both unipolar
and bipolar designs have been employed for many
years, a direct comparison of their performance has
not yet been possible. Unipolar devices have only
recently become available with head sizes that have l-
mm increments in diameter to allow for closer
matching of the head to the acetabular socket.
Moreover, unipolar devices have generally not been
made available
Correspondence to Dr M. Spector, Department of Orthopedic
Surgery, Brigham and Women’s Hospital, 75 Francis St.,
Boston, MA 02115, USA.
with ‘modern’
canal-filling
stem
designs. While direct comparison
of unipolar and
441
Biomaterials 1997,
Vol. 18 No. 5
4978778.002.png
442
Oxidized zirconium for hemiarthroplasty:
A.M. Pate/ and M. Spector
bipolar devices has not yet been possible, clinical
studies indicate that implementation of bipolar
prostheses has not significantly reduced acetabular
erosion”, while introducing concerns of polyethylene
wear. A recent studyI revealed femoral osteolysis in
58% of 45 primary uncemented bipolar hemiarthro-
plasties in 38 patients followed for an average of 5
years. Eight of these prostheses that were revised
displayed wear of the polyethylene insert rims without
substantial wear of the bearing surface. Histological
examination of the peri-implant tissue revealed
polyethylene particles in macrophages and
multinucleated foreign body giant cells. The authors
concluded that ‘the intrinsic functional mechanism of
the bipolar endoprosthesis leads to persistent cyclic
impingement of the stem neck on the rim of the
bipolar head, producing polyethylene particles and
frequent stem loosening by periprosthetic osteolysis’.
A critical element of the hemiarthroplasty prosthesis
is the material from which the head is produced and
its topography. Previous laboratory studies have
revealed some of the determinants of wear of articular
cartilage when it is articulated with metal counterfaces.
Lipshitz and Glimcher’” developed a laboratory test
apparatus to rub articular cartilage against a stainless
steel counterface. They found that the wear rate of
cartilage increased with roughness of the metal and
that the presence of a third body greatly increased
wear. In these experiments, calcium phosphate
particles were employed as the third bodies in order to
simulate the conditions when bone fragments or
mineral deposits are introduced between the
articulating components. Also of note was the finding
that there was no correlation between the orientation
of the cartilage plug, relative to the direction of sliding,
and wear rate.
Cobalt-chromium alloy is the primary material
employed as the bearing surface for the hemiarthro-
plasty device. However, animal14 as well as the
aforementioned human studies have documented the
degradative changes in articular cartilage associated
with the use of this material for hemiarthroplasty
prostheses. Moreover, comparable findings were
reported in an investigation employing cobalt-
chromium alloy devices to resurface the trochlea of the
dog knee’“. The study documented the degradative
changes in the articular cartilage of the opposing
patella 3 months after implantation of the hemiarthro-
plasty device. The investigation also demonstrated the
increase in wear rate with roughness of the metallic
surface.
In an attempt to develop prostheses with improved
tribological performance, several investigators have
studied the effects of the material of fabrication of the
hemiarthroplasty device on the wear of the opposing
articular cartilage using canine models for hemiarthro-
plasty of the hip. Mendes et al.lfi demonstrated the
unsuitability of high-density polyethylene as the
material of fabrication of the hemiarthroplasty head.
Significant wear of articular cartilage and the high-
density polyethylene was found with these devices.
Investigation of similar devices employed in human
studies yielded similar findings17. Canine studies
employing alumina heads identified some erosion
through the articular cartilage to bone after 2 years’a.
These studies demonstrated, however, that despite the
articulation of the head with bone, there was no
scratching of the ceramic. This finding led the investiga-
tors to recommend ceramic for hemiarthroplasty
devices. A more recent study showed that there was less
wear of articular cartilage with carbon heads than with
cobalt-chromium alloy or titanium alloy devices’“. In
summary, results of previous animal investigations have
suggested that some materials might be superior to
cobalt-chromium alloy for hemiarthroplasty. These
studies have confirmed laboratory experiments that
have shown that the wear rate of cartilage increases with
the roughness of the metal counterface. Benefits have
been demonstrated with the use of scratch-resistant
ceramic materials. Moreover, there has been some
evidence that certain materials, perhaps by virtue of
their lubricity, yield less wear of articular cartilage than
cobalt-chromium alloy.
The objective of this study was to investigate the
tribology of a novel abrasion-resistant material,
oxidized zirconium, when used as the counterface
against articular cartilage. An accelerated wear test
was used in this laboratory screen of the material for
hemiarthroplasty. Oxidized zirconium components are
produced by oxidizing particular zirconium alloys to
form a hard, abrasion-resistant ceramic surface’“.
The ceramic, zirconium oxide (zirconia), has proven
to be of value for decades for a wide range of non-
medical applications, and more recently for the
fabrication of femoral heads for total hip replacement
prostheses. In contrast, the metal, zirconium, and its
alloys owe their development principally to the
nuclear industry, where they have been able to meet
the special needs of a cladding material for pressurized
water vessels”. Zirconium alloys are commended as
nuclear core materials by their small neutron
macroscopic absorption cross-section”“. Advantages of
oxidized zirconium alloys for other applications, such
as the fabrication of components of total joint
replacement prostheses, relate to abrasion resistance
and thickness of the oxide layer. Transmission electron
microscopy
of the oxide scale on certain zirconium
alloys”2-2”
has provided evidence
of the adherent
nature of the oxide scale.
MATERIALS AND METHODS
Oxidized zirconium
Zirconium-2.5niobium alloy was oxidized using time
and temperature conditionszO to form an oxide 7pm in
thickness (Smith & Nephew Orthopaedics, Memphis,
TN, USA). Previous studies have demonstrated how
the thickness of the oxide layer and the size and
orientation of the blue-black monoclinic zirconium
oxide crystallites comprising the oxide can vary with
the oxidation conditions20,22. 23. 26. Discs fabricated
from oxidized zirconium and cast and wrought cobalt-
chromium alloy controls had a highly polished finish
(R, < 0.03 pm).
Laboratory apparatus for an accelerated wear
test
A pin-on-disc
wear apparatus
(Implant
Sciences
Company,
Danvers,
MA, USA) was modified
to
evaluate the tribological performance
of biomaterials
Biomaterials
1997, Vol. 18 No. 5
4978778.003.png
Oxidized zirconium for hemiarthroplasty:
A.M. Pate/ and M. Spector
443
with articular cartilage counterfaces. Osteochondral
plugs from calves and adult cows, 7mm in diameter,
served as the pin and the biomaterials as the disc. The
apparatus was set up so that the disc rotated in one
direction, with the pin circumscribing a path with a
radius of 7 mm (to the inner edge of the pin), yielding
22 736 revolutions per km. The test was configured to
maintain articular cartilage in a fluid environment at
room temperature while generating loaded sliding
contact with the discs. The apparatus was instrumented
with a Chatillon 250-g force transducer (Commercial
Scales) to measure the lateral load on the pin for
calculation of the coefficient of friction. A PC-based
data acquisition system was implemented for real-time
measurements of the friction forces.
Normal loads of 1.5 and 2.5 kg were used in this
study. We found that higher load, using the cobalt-
chromium alloy control in our test apparatus, resulted
in exceedingly high frictional forces and frequently in
shear failure of the articular cartilage from its bone
substrate. Assuming a contact area of approximately
0.3cmZ for the slightly convex cartilage specimen, the
contact stresses used in this study were about 5 and
8 kgcrn--’ for the 1.5 and 2.5 kg loads, respectively.
These values are at the low end of the physiological
rangez7, compared to the relatively high stress
(82 kg cm ‘) employed in a previous investigation13.
The articulation was lubricated with sterile, filtered
(triple filtration with 0.1 {trn filter) bovine calf serum
(Hyclone Labs, Inc.). McKellop and Luz8 had
previously reported a build-up of calcium phosphate
on the surface of zirconia balls tested in a hip simulator
with bovine serum as the lubricant. We performed X-
ray photoelectron spectroscopy on discs of oxidized
zirconium and cast cobalt-chromium
anomalous friction measurements due to edge effects.
The calf cartilage differed histologically from the adult
specimens in that the chondrocytes were less columnar
in their arrangement and the collagen less organized.
The histology of adult bovine articular cartilage is
similar to that of human material.
The bone portion of the plug was mounted in
poly(methy1 methacrylate) to facilitate stable fixation
to the wear test apparatus. Plugs were stored at 4°C in
phosphate-buffered
saline for a maximum of 5 days
until tested.
Coefficient of friction
Friction analysis was performed on oxidized
zirconium, cast cobalt-chromium and forged cobalt-
chromium articulated with the calf cartilage. Tests
were performed at speeds of under 0.1 ms-‘; this
speed was in the middle of the physiological range27.
The real-time measurements of the frictional forces
recorded during the tests were divided by the normal
load to obtain the coefficient of friction. These values
were plotted against slide distance. Coefficients of
friction for the materials were then determined by the
asymptotic values of the real-time measurements.
Statistical analysis of the difference in coefficient of
friction between groups was performed
using the
Student t-test.
Wear analysis
alloy specimens
Quantitative analysis of wear involved determination of
the linear wear by measuring the decrease in thickness
of the cartilage resulting from articulation with the test
and control counterface. Before a cartilage plug was
used in the wear test, the thickness of the cartilage was
measured with calipers at four points around the
circumference and then averaged to obtain an estimate
for the thickness of the cartilage at the centre of the
plug. Normal loads of 1.5 and 2.5 kg were applied and
tests were run for a slide distance of 9 km. Immediately
after testing, the plug was stored in 10% neutral
buffered formalin. After 24 h, the plug was sliced in
half, perpendicular to its surface, and the thickness at
the centre of the plug measured. This value was then
subtracted from the initial estimate to obtain the
change in thickness. Statistical analysis of the
difference in wear between groups was performed
using the Student t-test.
After taking the thickness measurement, one-half of the
cartilage was prepared for histological evaluation. The
cartilage was dehydrated in ethanol and embedded in
glycol methacrylate. Three-micrometre-thick sections
were cut with a Reichert-Jung microtome. These sections
were stained with Masson Trichrome for collagen,
Safranin O/Fast Green for proteoglycans, and Haematox-
ylin and Eosin. Histological evaluation of the specimens
served two purposes: it enabled validation of the wear
measurements and it provided an insight into the
mechanism by which the articular cartilage wears.
Three factors affecting the accuracy of the quantita-
tive analysis of linear wear were: (1) the accuracy of
the pretest estimate of the thickness of the cartilage, (2)
the amount the cartilage swelled due to the damage
that the collagen matrix incurred from the sliding
contact, and (3) the change in thickness of the cartilage
due to creep. We used five osteochondral
after testing. We found no indication
of a calcium
precipitate.
Therefore,
we did not employ chemical
additives
(namely,
chelating
agents)
to prevent
calcification in our lubricant.
The pin-on-disc test, with the operating conditions
used in this study, was employed as a screening
method that could be expected to accelerate the wear
of articular cartilage (compared to conditions in viva).
However, without knowledge of the wear mechanisms
associated with hemiarthroplasty prostheses in human
subjects, it was not possible to determine the degree to
which the wear process was accelerated. That the pin-
on-disc apparatus did not closely simulate in vivo
conditions did not preclude its use for the comparison
of the tribological performance of a new material with
the clinical standard (i.e. cobalt-chromium
alloy).
Osteochondral plug preparation
Stifle joints of calves and adult cows were obtained
within 4 h of slaughter (Research 87, Hopkinton, MA,
USA). The surface of the femoral condyles was isolated
and kept moist with a solution of phosphate-buffered
saline. The technique used for preparation of the
osteochondral plugs is described elsewherezg. The
cartilage on the plugs had a diameter of 7mm and
ranged in thickness from 1.0 to 2.5mm for the adult
cartilage and 4 to 7 mm for the calf samples. All of the
plugs were cored from the same region of the femoral
condyles. Osteochondral plugs from this site had a
slightly convex surface. This geometry
minimized
plugs to
Biomaterials 1997, Vol. 18 No. 5
4978778.004.png
444
Oxidized zirconium for hemiarthroplasty: A.M. Pate/ and M. Spector
evaluate the accuracy of the pretest estimate. Four
circumferential measurements of the thickness of the
cartilage were taken for each plug. The plugs were
then sliced perpendicular to their surfaces and
thickness measurements at the centre of the plugs
were taken. These values were then compared with the
average value of the circumferential measurements to
evaluate the accuracy of the pretest estimate. There
was no significant difference. The degree to which
swelling and creep could affect the determination of
linear wear was assessed in experiments using control
samples of calf cartilagezg.
After wear testing, the test and control counterfaces
were gently rinsed with water to remove any residue
of the bovine serum lubricant. Samples were sputter
coated with gold. The counterfaces were then evaluated
using conventional scanning electron microscopy.
wear layer from the underlying cartilage, which
appeared normal with respect to its staining for collagen
and proteoglycan. The thickness, stain characteristics
and consistency of the wear layer produced by articula-
tion with the two types of biomaterials were comparable.
On some specimens, the wear layer was non-uniform in
thickness, accumulating on one side of the plug (Figure
3), apparently related to the direction of rotation of the
disc. When stained with Safranin O/Fast Green, the
RESULTS
Swelling of specimens of cartilage due to the damaged
collagen, produced by removing the superficial layer
of control specimens of cartilage with a scalpel, was
found to be 0.1 f 0.1 mm (mean* s.d.; n = 6). The
effect of creep, under the constant load of 2.5 kg used
in the wear experiment, was to decrease the thickness
of the cartilage 0.1 f O.Omm (n = 4). Based on these
findings, it was presumed that there was no remarkable
net effect of creep and swelling on the linear wear
measurements.
The linear wear of the adult bovine cartilage
articulated with the oxidized zirconium was
approximately 30% less than the wear of cartilage tested
with the cast cobalt-chromium alloy counterface
(0.34 f 0.13 mm versus 0.47 f 0.20 mm, mean f sd.;
Figure 1). This difference was, however, not statistically
significant (P > 0.1). A distinct layer of material, which
presumably was degraded cartilage, was found on the
surface of all of the osteochondral plugs after articulation
with the test and control materials. This ‘wear layer’ was
generally uniform in histological appearance and
thickness (approximately 50-75 pm) across the surface
(Figure 2). There was a distinct border separating the
T 1
1
0.5
0.4 1
T 1
Linear
I
Wear
(mm)
0.3
0.2
0.1
I
OJ---
ii&z
Ox kc
n=6
n=6
Figure 2 a, Micrograph of adult bovine cartilage stained
with haematoxylin and eosin (H&E). b, Section of adult
bovine cartilage that has articulated against oxidized
zirconium for 9 km. A 50-pm-thick wear layer is apparent.
H&E staining. c, Adult bovine cartilage that has articulated
against cast cobalt-chromium
Figure 1 Graph showing the linear wear of articular
cartilage with oxidized zirconium and cast cobalt-
chromium counterfaces. The heights of the columns
correspond to the mean values and the vertical bars the
standard errors of the mean.
for 9 km. H&E staining.
Biomaterials 1997, Vol. 18 No. 5
4978778.005.png
Oxidized zirconium for hemiarthroolasty:
A.M. Pate/ and M. Spector
445
0.0Z5
E
T
r 1
Coefficient
0.015
of
Friction
0.01
Castr Forged CoCr
n=4
-
Ox Zirc
II=4
n=4
Figure 3 Adult bovine cartilage that articulated with cast
cobalt-chromium for 9 km. A wear layer (200-250pm thick)
accumulated on one side of the plug, apparently associated
with the direction of rotation of the disc (i.e. from left to
right). H&E staining.
Figure 4 Comparison of the coefficients of friction for
articular cartilage articulated with oxidized zirconium and
cobalt-chromium alloy specimens. The heights of the
columns correspond to the mean values and the vertical
bars the standard errors of the mean.
degraded cartilage appeared to be deficient in
proteoglycan. This loss of proteoglycan was consistent
with a failure mechanism found in an animal model of
hemiarthroplasty14.
One of four of the calf cartilage plugs articulating
with the oxidized zirconium and three of four
samples in the groups articulating with the cast and
forged cobalt-chromium samples failed prior to
completion of the test. The articular cartilage sheared
from the subchondral bone. This appeared to be
related to elevations in the frictional forcezg. The
wear for the remaining cartilage samples tested
against the oxidized zirconium was 0.5 f 0.1 mm
(meanfs.d.; n = 3) compared to 0.8mm (n = 1) for
the plugs with the forged counterface and 1 mm for
the cartilage
a
with the cast cobalt-chromium
specimenZg.
Using the 1.5 kg load in tests with the calf cartilage,
the coefficient of friction for the oxidized zirconium
was 0.015 f 0.002 (meanfs.d.; R = 4; Figure 4). This
was significantly less (P < 0.0001) than the values
for the cast (0.021 f0.003; n = 4) and the forged
(0.023 f 0.003; n = 4) cobalt-chromium specimens.
Results from tests of adult bovine cartilage using the
2.5 kg load were variable. In half of the tests of the
oxidized zirconium and cast cobalt-chromium
counterfaces, the coefficients of friction were initially
approximately 0.01, rising to more than 0.17; there
was an abrupt increase in frictional force that
exceeded the force transducer’s capacity. The other
half of the tests started with frictional forces that
were out of range of the transducer. Only one of the
tests with cast cobalt-chromium alloy returned to
range, whereas two of the oxidized zirconium
returned.
Wear tracks were apparent on all of the counterfaces.
On certain discs, there was a clear demarcation
between regions that appeared to have a transfer layer
of degraded cartilage and the featureless regions where
articulation had not occurred. On other counterfaces,
there was a gradual decrease of the amount of transfer
film as distance from the centre of the wear track
increased (Figure 5). There were no marked differences
in structure of the transfer layer on the test and control
counterfaces.
b
Figure 5 a, Scanning electron micrograph of the centre of
a wear track formed on a disc of cast cobalt-chromium
alloy, displaying a transfer layer of articular cartilage from
the osteochondral plug. b, Scanning electron micrograph
showing the transfer layer on a cast cobalt-chromium
alloy
disc at approximately
10pm from the centre of a wear track.
DISCUSSION
Results of this study showed that the coefficient of
friction of articular cartilage rubbed against an oxidized
zirconium counterface was significantly lower than
Biomaterials
1997, Vol. 18 No. 5
T
4978778.001.png
Zgłoś jeśli naruszono regulamin